2.8 KiB
2.8 KiB
Sinara 7210 Clocker
JSON
Not present in the JSON.
Peripherals typically should choose "clk_sel": 2
for MMCX connection and "clk_sel": 1
for external SMA connection.
Refer to the official docs by searching for clk_sel
.
You may also need to add "refclk": <number>
field to the target card.
Setup external clocker
For tests, you may need an external RF generator, depending on customer needs. Here is example setup for SynthNV RF signal generator:
- Connect SynthNV to the workstation via USB, and
- Install and run
cutecom
:nix-shell -p cutecom
- Set settings as on the picture below:
- Open the device, usually it is
/dev/ttyACM0
- Put
?
intoInput
field and pressEnter
for current settings and help commands - For changing the frequency, enter
f<freq in MHz>
, e.g.f125.0
for 125 MHz - Set RF power so that clocker would recognize the signal with
a<power>
command, e.g.a63
- Check for desired amplitude and frequency at the
RFOut
(see picture below for reference) pin via oscilloscope - If everything is ok, connect
RFOut
to theCLK IN
on the Clocker (see instructions below for details)
Setup the Clocker
- Switch
CLK SEL
pin toEXT
/INT
according to customer needs - Connect MMCx cables according to the customer needs and boards specifications (see image below for reference):
if the
INT
source is chosen, connect MMCx cable toINT CLK
, otherwise connect external clocker to SMAEXT CLK
- Connect the Clocker to the Kasli via 30-pin ports, or via external power supply
- Connect the Clocker's SMA output to the Kasli's
CLK
/CLK IN
SMA pin - After assembling the crates and flashing the firmware, start Kasli and set config if needed:
artiq_coremgmt config write -s rtio_clock ext0_bypass
. Please refer to the official manual for the details and available options. In most cases you may skip this step. - Reboot either via
artiq_coremgmt reboot
or via power supply if the board's firmware doesn't have such command
Testing
Run artiq_sinara_test
and check that it doesn't fail on the connected devices.
Alternatively, if it would be shipped standalone:
- Switch to external source
- Connect to the external
CLK IN
clock source (frequency generator) via SMA cable - Power up Clocker with power supply or EEM
- Check via oscilloscope all (internal and external) clocker outputs, that they output clock signal respective to the input frequency
- Shut down Clocker
- Switch to internal source
- Connect clock source to the internal
CLK IN