linuswck
|
0a992c1dc8
|
ld: pd_responsitivity -> pd_mon_params
|
2024-04-10 16:57:43 +08:00 |
linuswck
|
489333c43d
|
ld: save pwr_limit into settings
|
2024-04-10 16:56:20 +08:00 |
linuswck
|
7d41edef20
|
ld: Increase the delay before turning on LD_PWR
- Delay is tested and verified with oscilloscope
- Tested with 0.3A LD output current
|
2024-03-21 15:21:24 +08:00 |
linuswck
|
607da8a822
|
ld: Add a delay before turning on LD PWR
- DAC output might not drop to 0V before PWR is being turned on
- Thus, it might drive nondeterministic current into the laser diode during pwr up
|
2024-03-20 17:24:47 +08:00 |
linuswck
|
dbbd438e92
|
ld: Fix unable to save & load ld_i flash settings
|
2024-03-19 14:48:25 +08:00 |
linuswck
|
3ad7c417f6
|
Fix a ld current ramp up bug
- During ld_i ramp up, if pwr up cmd is called, ld_i is set incorrectly
|
2024-03-18 16:41:55 +08:00 |
linuswck
|
111d9a4226
|
cleanup
|
2024-03-05 16:33:49 +08:00 |
linuswck
|
92c6cf12d4
|
Add fns to apply ld, tec settings summary configs
|
2024-03-05 10:35:05 +08:00 |
linuswck
|
ef2410f441
|
Add fns, struct items for getting ld, tec settings
|
2024-03-05 10:35:05 +08:00 |
linuswck
|
f488786e1c
|
Use DMA to read Tec_I & Tec_V
- Remove the single conversion polling time with DMA
|
2024-03-01 11:31:01 +08:00 |
linuswck
|
76477065be
|
Change f64 to f32 for SI Unit related computation
- Save computational time
- Improve the processing speed of TecSetI cmd
|
2024-02-28 12:59:58 +08:00 |
linuswck
|
18dd0a7963
|
Save Thermostat and Laser Diode PWR On/Off status
|
2024-02-28 12:22:40 +08:00 |
linuswck
|
c09ccc29cd
|
cmd: Add cmd to get all device settings
|
2024-02-23 12:45:29 +08:00 |
linuswck
|
c4135f6ac3
|
cmd: Add cmds to config pd_mon params
|
2024-02-22 16:44:45 +08:00 |
linuswck
|
ed785b7c85
|
ld: correct timer and dac reset in pwr_up seq
|
2024-02-21 11:11:12 +08:00 |
linuswck
|
30ab228b4b
|
ld: Set default ld_current_limit in setup() fn
|
2024-02-20 16:22:13 +08:00 |
linuswck
|
88cca12a60
|
ld: Use Timer IRQ to ramp up/down output current
- Fixes possible watchdog timeout
- Improves ethernet response time & thermostat Pid update interval consistency
|
2024-02-20 16:18:06 +08:00 |
linuswck
|
5f83b73011
|
cmd: Add active report mode cmd
|
2024-02-19 15:08:00 +08:00 |
linuswck
|
60a79d1780
|
rename get_term_status -> get_lf_mod_in_impedance
|
2024-02-15 12:14:14 +08:00 |
linuswck
|
6096711d2c
|
eth: Poll and Handle Json based Ld ctrl cmd
- Upgrade to miniconf 9.0
- Only non report related laser diode ctrl command is implemented
|
2024-02-02 17:35:58 +08:00 |
linuswck
|
3d3d6f5cb5
|
ld: cleanup
|
2024-02-02 17:23:39 +08:00 |
linuswck
|
4e0d14def2
|
laser: Add driver and fn to get termination status
|
2024-01-30 13:01:33 +08:00 |
linuswck
|
f6677d874c
|
LD: Rst LD current at power_up for soft start
|
2024-01-26 16:56:55 +08:00 |
linuswck
|
d391e3a1fb
|
LD: Setting LD output current now ramps up slowly
|
2024-01-26 16:56:55 +08:00 |
linuswck
|
edd30e94a0
|
laser_diode: Add fn to set pd_dark_i
|
2024-01-26 16:56:55 +08:00 |
linuswck
|
ad731c2f15
|
rename pd_mon -> pd_responsitivity
|
2024-01-26 16:56:55 +08:00 |
linuswck
|
9d8a553669
|
pd_mon: add fns to convert between pwr and current
|
2024-01-26 16:56:55 +08:00 |
linuswck
|
f50505feaf
|
Rename Analog_Wdg -> LdPwrExcProtector
|
2024-01-26 16:56:55 +08:00 |
linuswck
|
85b50bf824
|
Analog_Wdg: Finish Power Excursion Protection fns
- Verified to be working
|
2024-01-26 16:56:39 +08:00 |
linuswck
|
af283b17ac
|
laser_diode: use analog wdg fns
|
2024-01-24 12:18:42 +08:00 |
linuswck
|
0d12c902fc
|
pd_mon: Separate ld_power calculation to a file
|
2024-01-24 12:18:37 +08:00 |
linuswck
|
3cfdee917a
|
rename current_source -> ld_ctrl
|
2024-01-24 12:16:06 +08:00 |
linuswck
|
5f582be143
|
rename ld_drive to laser_diode
|
2024-01-24 12:16:06 +08:00 |