LD overcurrent with all modulation switches are OFF #35
Loading…
Reference in New Issue
Block a user
No description provided.
Delete Branch "%!s()"
Deleting a branch is permanent. Although the deleted branch may continue to exist for a short time before it actually gets removed, it CANNOT be undone in most cases. Continue?
Not selecting any modulation depth (all switches to OFF) results in an additional 120mA of laser current.
This represents a non-obvious way of damaging the laser diode and should be fixed in the next revision.