Clocker "rise time" spec is likely incorrect #44

Closed
opened 2022-07-02 18:05:11 +08:00 by sb10q · 2 comments
Owner

The datasheet says "max frequency 4000MHz". That's a 250ps period. The 415ps rise time spec is contradictory.

The datasheet says "max frequency 4000MHz". That's a 250ps period. The 415ps rise time spec is contradictory.
Author
Owner

If this was measured with the Siglent scope, I don't think it's fast enough for that.

If this was measured with the Siglent scope, I don't think it's fast enough for that.
Member

Removed in a14aa89.
The rise time from -400 mV to +400 mV is estimated to be ~100 ps in the testing thread. But there aren't any good plot there (except a blurry plot with the cursors not at the +-400 mV setpoints).

Removed in a14aa89. The rise time from -400 mV to +400 mV is estimated to be ~100 ps in the testing thread. But there aren't any good plot there (except a blurry plot with the cursors not at the +-400 mV setpoints).