document Urukul nonlinearity #41

Closed
opened 2022-06-23 11:37:17 +08:00 by sb10q · 4 comments
https://forum.m-labs.hk/d/445-nonlinear-urukul-dds-amplitude
Collaborator

image
This looks nowhere as bad as the plot on the forum, even with no attenuation & ASF=1.

![image](/attachments/5d6b97ff-fd62-4cbd-915b-0c120e597342) This looks nowhere as bad as the plot on the forum, even with no attenuation & ASF=1.
Poster
Owner

How was this data obtained?

How was this data obtained?
Collaborator

Urukul 4410 direct attached to an oscilloscope with 50 Ohm termination in the scope. Value obtained from the scope.

Urukul 4410 direct attached to an oscilloscope with 50 Ohm termination in the scope. Value obtained from the scope.
Collaborator

6d9faa7b
Added plots to demonstrate the impact of attenuation on linearity.
Chose 15dB attenuation just for consistency with the SU-Servo example, which chose 15dB attenuation to maintain linearity.

image

6d9faa7b Added plots to demonstrate the impact of attenuation on linearity. Chose 15dB attenuation just for consistency with the SU-Servo example, which chose 15dB attenuation to maintain linearity. ![image](/attachments/c19928d9-4bdb-454c-b9bb-d0a2507da195)
118 KiB