add contact email

This commit is contained in:
Sebastien Bourdeauducq 2019-11-28 11:19:03 +08:00
parent a66e9b143c
commit 24b3b08f7c
1 changed files with 2 additions and 0 deletions

View File

@ -77,6 +77,8 @@ Current status
* Design may contain errors and things that can be simplified or optimized further. This is a highly experimental system and YMMV.
* HP 5501B lasers do not like light (at any wavelength) sent into their aperture; the stabilization circuit fails and the REF output becomes wrong. Workaround is to misalign the input beam, which obviously introduces cosine error. A better solution is needed, maybe rebuild a 2-frequency HeNe but use the waste beam of the tube for REF/intensity measurements unlike the HP design. The HeNe tube itself attenuates incoming light.
Questions, comments: email sb [AT] m-labs.hk
Software
--------