thermostat/pytec/pytec/client.py
2020-10-01 00:35:16 +02:00

164 lines
5.1 KiB
Python

import socket
import json
CHANNELS = 2
class Client:
def __init__(self, host="192.168.1.26", port=23, timeout=None):
self._socket = socket.create_connection((host, port), timeout)
self._lines = [""]
def _command(self, *command):
self._socket.sendall((" ".join(command) + "\n").encode('utf-8'))
def _read_line(self):
# read more lines
while len(self._lines) <= 1:
chunk = self._socket.recv(4096)
if not chunk:
return None
buf = self._lines[-1] + chunk.decode('utf-8', errors='ignore')
self._lines = buf.split("\n")
line = self._lines[0]
self._lines = self._lines[1:]
return line
def _get_conf(self, topic):
self._command(topic)
result = []
for channel in range(0, CHANNELS):
line = self._read_line()
conf = json.loads(line)
result.append(conf)
return result
def get_pwm(self):
"""Retrieve PWM limits for the TEC
Example::
[{'channel': 0,
'center': 'vref',
'i_set': {'max': 2.9802790335151985, 'value': -0.02002179650216762},
'max_i_neg': {'max': 3.0, 'value': 3.0},
'max_v': {'max': 5.988, 'value': 5.988},
'max_i_pos': {'max': 3.0, 'value': 3.0}},
{'channel': 1,
'center': 'vref',
'i_set': {'max': 2.9802790335151985, 'value': -0.02002179650216762},
'max_i_neg': {'max': 3.0, 'value': 3.0},
'max_v': {'max': 5.988, 'value': 5.988},
'max_i_pos': {'max': 3.0, 'value': 3.0}}
]
"""
return self._get_conf("pwm")
def get_pid(self):
"""Retrieve PID control state
Example::
[{'channel': 0,
'parameters': {
'kp': 10.0,
'ki': 0.02,
'kd': 0.0,
'output_min': 0.0,
'output_max': 3.0,
'integral_min': -100.0,
'integral_max': 100.0},
'target': 37.0,
'integral': 38.41138597026372},
{'channel': 1,
'parameters': {
'kp': 10.0,
'ki': 0.02,
'kd': 0.0,
'output_min': 0.0,
'output_max': 3.0,
'integral_min': -100.0,
'integral_max': 100.0},
'target': 36.5,
'integral': nan}]
"""
return self._get_conf("pid")
def get_steinhart_hart(self):
"""Retrieve Steinhart-Hart parameters for resistance to temperature conversion
Example::
[{'params': {'b': 3800.0, 'r0': 10000.0, 't0': 298.15}, 'channel': 0},
{'params': {'b': 3800.0, 'r0': 10000.0, 't0': 298.15}, 'channel': 1}]
"""
return self._get_conf("s-h")
def get_postfilter(self):
"""Retrieve DAC postfilter configuration
Example::
[{'rate': None, 'channel': 0},
{'rate': 21.25, 'channel': 1}]
"""
return self._get_conf("postfilter")
def report_mode(self):
"""Start reporting measurement values
Example of yielded data::
{'channel': 0,
'time': 2302524,
'adc': 0.6199188965423515,
'sens': 6138.519310282602,
'temperature': 36.87032392655527,
'pid_engaged': True,
'i_set': 2.0635816680889123,
'vref': 1.494,
'dac_value': 2.527790834044456,
'dac_feedback': 2.523,
'i_tec': 2.331,
'tec_i': 2.0925,
'tec_u_meas': 2.5340000000000003,
'pid_output': 2.067581958092247}
"""
self._command("report mode", "on")
while True:
line = self._read_line()
if not line:
break
try:
yield json.loads(line)
except json.decoder.JSONDecodeError:
pass
def set_param(self, topic, channel, field="", value=""):
"""Set configuration parameters
Examples::
tec.set_param("pwm", 0, "max_v", 2.0)
tec.set_param("pid", 1, "output_max", 2.5)
tec.set_param("s-h", 0, "t0", 20.0)
tec.set_param("center", 0, "vref")
tec.set_param("postfilter", 1, 21)
See the firmware's README.md for a full list.
"""
if type(value) is float:
value = "{:f}".format(value)
if type(value) is not str:
value = str(value)
self._command(topic, str(channel), field, value)
def power_up(self, channel, target):
"""Start closed-loop mode"""
self.set_param("pid", channel, "target", value=target)
self.set_param("pwm", channel, "pid")
def save_config(self):
"""Save current configuration to EEPROM"""
self._command("save")
def load_config(self):
"""Load current configuration from EEPROM"""
self._command("load")