@ -88,3 +88,94 @@ with logging via semihosting.)
|
|||
**Caveat:** This logging does not flush its output. Doing so would
|
||||
hang indefinitely if the output is not read by the USB host. Therefore
|
||||
output will be truncated once buffers are full.
|
||||
|
||||
|
||||
## Temperature measurement
|
||||
|
||||
Connect the thermistor with the SENS pins of the
|
||||
device. Temperature-depending resistance is measured by the AD7172
|
||||
ADC. To prepare conversion to a temperature, set the Beta parameters
|
||||
for the Steinhart-Hart equation.
|
||||
|
||||
Set the base temperature in degrees celsius for the channel 0 thermistor:
|
||||
```
|
||||
s-h 0 t0 20
|
||||
```
|
||||
|
||||
Set the resistance in Ohms measured at the base temperature t0:
|
||||
```
|
||||
s-h 0 r0 10000
|
||||
```
|
||||
|
||||
Set the Beta parameter:
|
||||
```
|
||||
s-h 0 b 3800
|
||||
```
|
||||
|
||||
|
||||
## Thermo-Electric Cooling (TEC)
|
||||
|
||||
### Limits
|
||||
|
||||
Each of the MAX1968 TEC driver has analog/PWM inputs for setting
|
||||
output limits.
|
||||
|
||||
Use the `pwm` command to see current settings and maximum values.
|
||||
|
||||
| Limit | Unit | Description |
|
||||
| --- | :---: | --- |
|
||||
| max_v | Volts | Maximum voltage |
|
||||
| max_i_pos | Amperes | Maximum positive current |
|
||||
| max_i_neg | Amperes | Maximum negative current |
|
||||
| | Amperes | current control |
|
||||
|
||||
Example: set the maximum voltage of channel 0 to 1.5 V.
|
||||
```
|
||||
pwm 0 max_v 1.5
|
||||
```
|
||||
|
||||
### Open-loop mode
|
||||
|
||||
To manually control TEC output current, omit the limit parameter of
|
||||
the `pwm` command. Doing so will disengage the PID control for that
|
||||
channel.
|
||||
|
||||
Example: set output current of channel 0 to 0 A.
|
||||
```
|
||||
pwm 0 0
|
||||
```
|
||||
|
||||
## PID-stabilized temperature control
|
||||
|
||||
Set the target temperature of channel 0 to 20 degrees celsius:
|
||||
```
|
||||
pid 0 target 20
|
||||
```
|
||||
|
||||
Enter closed-loop mode by switching control of the TEC output current
|
||||
of channel 0 to the PID algorithm:
|
||||
```
|
||||
pwm 0 pid
|
||||
```
|
||||
|
||||
## Reports
|
||||
|
||||
Use the bare `report` command to obtain a single report. Enable
|
||||
continuous reporting with `report mode on`. Reports are JSON objects
|
||||
with the following keys.
|
||||
| Key | Unit | Description |
|
||||
| --- | :---: | --- |
|
||||
| `channel` | Integer | Channel `0`, or `1` |
|
||||
| `time` | Milliseconds | Temperature measurement time |
|
||||
| `adc` | Volts | AD7172 input |
|
||||
| `sens` | Ohms | Thermistor resistance derived from `adc` |
|
||||
| `temperature` | Degrees Celsius | Steinhart-Hart conversion result derived from `sens` |
|
||||
| `pid_engaged` | Boolean | `true` if in closed-loop mode |
|
||||
| `i_set` | Amperes | TEC output current |
|
||||
| `vref` | Volts | MAX1968 VREF (1.5 V) |
|
||||
| `dac_value` | Volts | AD5680 output derived from `i_set` |
|
||||
| `dac_feedback` | Volts | ADC measurement of the AD5680 output |
|
||||
| `i_tec` | Volts | MAX1968 TEC current monitor |
|
||||
| `tec_i` | Amperes | TEC output current feedback derived from `i_tec` |
|
||||
| `tec_u_meas` | Volts | Measurement of the voltage across the TEC |
|
||||
| `pid_output` | Amperes | PID control output |
|
||||
|
|