ionpak-thermostat/firmware/src/main.rs

218 lines
6.4 KiB
Rust

#![feature(used, const_fn, core_float)]
#![no_std]
extern crate cortex_m;
extern crate cortex_m_rt;
extern crate tm4c129x;
use core::cell::{Cell, RefCell};
use core::fmt;
use cortex_m::exception::Handlers as ExceptionHandlers;
use cortex_m::interrupt::Mutex;
use tm4c129x::interrupt::Interrupt;
use tm4c129x::interrupt::Handlers as InterruptHandlers;
#[macro_export]
macro_rules! print {
($($arg:tt)*) => ({
use core::fmt::Write;
write!($crate::UART0, $($arg)*).unwrap()
})
}
#[macro_export]
macro_rules! println {
($fmt:expr) => (print!(concat!($fmt, "\n")));
($fmt:expr, $($arg:tt)*) => (print!(concat!($fmt, "\n"), $($arg)*));
}
#[macro_use]
mod board;
mod pid;
mod loop_anode;
mod loop_cathode;
mod electrometer;
static TIME: Mutex<Cell<u64>> = Mutex::new(Cell::new(0));
fn get_time() -> u64 {
cortex_m::interrupt::free(|cs| {
TIME.borrow(cs).get()
})
}
static LOOP_ANODE: Mutex<RefCell<loop_anode::Controller>> = Mutex::new(RefCell::new(
loop_anode::Controller::new()));
static LOOP_CATHODE: Mutex<RefCell<loop_cathode::Controller>> = Mutex::new(RefCell::new(
loop_cathode::Controller::new()));
static ELECTROMETER: Mutex<RefCell<electrometer::Electrometer>> = Mutex::new(RefCell::new(
electrometer::Electrometer::new()));
pub struct UART0;
impl fmt::Write for UART0 {
fn write_str(&mut self, s: &str) -> Result<(), fmt::Error> {
for c in s.bytes() {
unsafe {
let uart_0 = tm4c129x::UART0.get();
while (*uart_0).fr.read().txff().bit() {}
(*uart_0).dr.write(|w| w.data().bits(c))
}
}
Ok(())
}
}
fn main() {
board::init();
cortex_m::interrupt::free(|cs| {
// Enable FPU
let scb = tm4c129x::SCB.borrow(cs);
scb.enable_fpu();
let nvic = tm4c129x::NVIC.borrow(cs);
nvic.enable(Interrupt::ADC0SS0);
let mut loop_anode = LOOP_ANODE.borrow(cs).borrow_mut();
let mut loop_cathode = LOOP_CATHODE.borrow(cs).borrow_mut();
/*
// ZJ-10
let anode = 165.0;
let cathode_bias = 50.0;
let emission = 0.5e-3;
*/
/*
// ZJ-27
let anode = 225.0;
let cathode_bias = 25.0;
let emission = 1.0e-3;
*/
// ZJ-12
/*let anode = 200.0;
let cathode_bias = 50.0;
let emission = 4.0e-3;*/
let anode = 180.0;
let cathode_bias = 30.0;
let emission = 4.0e-3;
loop_anode.set_target(anode);
loop_cathode.set_emission_target(emission);
loop_cathode.set_bias_target(cathode_bias);
});
println!(r#"
_ _
(_) | |
_ ___ _ __ _ __ __ _| |
| |/ _ \| '_ \| '_ \ / _` | |/ /
| | (_) | | | | |_) | (_| | <
|_|\___/|_| |_| .__/ \__,_|_|\_\
| |
|_|
Ready."#);
let mut next_blink = 0;
let mut next_info = 0;
let mut led_state = true;
let mut latch_reset_time = None;
loop {
board::process_errors();
let time = get_time();
if time > next_blink {
led_state = !led_state;
next_blink = time + 100;
board::set_led(1, led_state);
}
if time > next_info {
// FIXME: done in ISR now because of FPU snafu
/*cortex_m::interrupt::free(|cs| {
LOOP_CATHODE.borrow(cs).borrow().debug_print();
});*/
next_info = next_info + 300;
}
if board::error_latched() {
match latch_reset_time {
None => {
println!("Protection latched");
latch_reset_time = Some(time + 1000);
}
Some(t) => if time > t {
latch_reset_time = None;
cortex_m::interrupt::free(|cs| {
// reset PID loops as they have accumulated large errors
// while the protection was active, which would cause
// unnecessary overshoots.
LOOP_ANODE.borrow(cs).borrow_mut().reset();
LOOP_CATHODE.borrow(cs).borrow_mut().reset();
board::reset_error();
});
println!("Protection reset");
}
}
}
}
}
use tm4c129x::interrupt::ADC0SS0;
extern fn adc0_ss0(_ctxt: ADC0SS0) {
cortex_m::interrupt::free(|cs| {
let adc0 = tm4c129x::ADC0.borrow(cs);
if adc0.ostat.read().ov0().bit() {
panic!("ADC FIFO overflowed")
}
adc0.isc.write(|w| w.in0().bit(true));
let ic_sample = adc0.ssfifo0.read().data().bits();
let fbi_sample = adc0.ssfifo0.read().data().bits();
let fv_sample = adc0.ssfifo0.read().data().bits();
let fd_sample = adc0.ssfifo0.read().data().bits();
let av_sample = adc0.ssfifo0.read().data().bits();
let fbv_sample = adc0.ssfifo0.read().data().bits();
let mut loop_anode = LOOP_ANODE.borrow(cs).borrow_mut();
let mut loop_cathode = LOOP_CATHODE.borrow(cs).borrow_mut();
let mut electrometer = ELECTROMETER.borrow(cs).borrow_mut();
loop_anode.adc_input(av_sample);
loop_cathode.adc_input(fbi_sample, fd_sample, fv_sample, fbv_sample);
electrometer.adc_input(ic_sample);
let time = TIME.borrow(cs);
time.set(time.get() + 1);
if time.get() % 300 == 0 {
println!("");
/*loop_anode.get_status().debug_print();
loop_cathode.get_status().debug_print();
electrometer.get_status().debug_print();*/
let cathode_status = loop_cathode.get_status();
let electrometer_status = electrometer.get_status();
if cathode_status.fbi.is_some() && electrometer_status.ic.is_some() {
let fbi = cathode_status.fbi.unwrap();
let ic = electrometer_status.ic.unwrap();
let pressure = ic/fbi/18.75154;
println!("{:.1e} mbar", pressure);
}
}
});
}
#[used]
#[link_section = ".rodata.exceptions"]
pub static EXCEPTIONS: ExceptionHandlers = ExceptionHandlers {
..cortex_m::exception::DEFAULT_HANDLERS
};
#[used]
#[link_section = ".rodata.interrupts"]
pub static INTERRUPTS: InterruptHandlers = InterruptHandlers {
ADC0SS0: adc0_ss0,