# Kasli_SOC
|
|
|
|
## HW Setup
|
|
|
|
Check the BOOT mode switches - they both should be at SD if the Kasli-SoC going to be shipped to customer.
|
|
POR jumper needs only for JTAG mode.
|
|
|
|
![](../img/kasli_soc.jpg)
|
|
|
|
## Mounting fan onto heatsink
|
|
![](../img/kasli_soc_fan.jpg)
|
|
1. ⚠️ Verify the fan has the **correct polarity (powering on with wrong polarity will burn the MOSFET in series💥)**
|
|
2. Place the fan on a heatsink
|
|
3. Tap 3 threads on the heatsink using M2.5 pointy tapping screws (e.g. front panel screws)
|
|
4. Replace the tapping screws with M2.5x14mm screws
|
|
5. Verify the fan is secure
|