README.md
11
README.md
|
@ -23,7 +23,7 @@ cargo build --release
|
||||||
|
|
||||||
The resulting ELF file will be located under `target/thumbv7em-none-eabihf/release/thermostat`.
|
The resulting ELF file will be located under `target/thumbv7em-none-eabihf/release/thermostat`.
|
||||||
|
|
||||||
Alternatively, you can install the Rust toolchain without Nix using rustup; see the Rust manifest file pulled in `flake.nix` to determine which Rust version to use.
|
Alternatively, you can install the Rust toolchain without Nix using rustup; see the `rust` variable in `flake.nix` to determine which Rust version to use.
|
||||||
|
|
||||||
## Debugging
|
## Debugging
|
||||||
|
|
||||||
|
@ -94,7 +94,7 @@ formatted as line-delimited JSON.
|
||||||
|
|
||||||
| Syntax | Function |
|
| Syntax | Function |
|
||||||
|------------------------------------------- |-------------------------------------------------------------------------------|
|
|------------------------------------------- |-------------------------------------------------------------------------------|
|
||||||
| `report` | Show current input |
|
| `report` | Show latest report of channel parameters (see *Reports* section) |
|
||||||
| `pwm` | Show current PWM settings |
|
| `pwm` | Show current PWM settings |
|
||||||
| `pwm <0/1> max_i_pos <amp>` | Set maximum positive output current, clamped to [0, 2] |
|
| `pwm <0/1> max_i_pos <amp>` | Set maximum positive output current, clamped to [0, 2] |
|
||||||
| `pwm <0/1> max_i_neg <amp>` | Set maximum negative output current, clamped to [0, 2] |
|
| `pwm <0/1> max_i_neg <amp>` | Set maximum negative output current, clamped to [0, 2] |
|
||||||
|
@ -189,7 +189,7 @@ Testing heat flow direction with a low set current is recommended before install
|
||||||
|
|
||||||
### Limits
|
### Limits
|
||||||
|
|
||||||
Each of the MAX1968 TEC driver has analog/PWM inputs for setting
|
Each MAX1968 TEC driver has analog/PWM inputs for setting
|
||||||
output limits.
|
output limits.
|
||||||
|
|
||||||
Use the `pwm` command to see current settings and maximum values.
|
Use the `pwm` command to see current settings and maximum values.
|
||||||
|
@ -218,7 +218,7 @@ pwm 0 max_i_pos 3
|
||||||
|
|
||||||
### Open-loop mode
|
### Open-loop mode
|
||||||
|
|
||||||
To manually control TEC output current, omit the limit parameter of
|
To manually control TEC output current, set a fixed output current with
|
||||||
the `pwm` command. Doing so will disengage the PID control for that
|
the `pwm` command. Doing so will disengage the PID control for that
|
||||||
channel.
|
channel.
|
||||||
|
|
||||||
|
@ -257,6 +257,7 @@ with the following keys.
|
||||||
| --- | :---: | --- |
|
| --- | :---: | --- |
|
||||||
| `channel` | Integer | Channel `0`, or `1` |
|
| `channel` | Integer | Channel `0`, or `1` |
|
||||||
| `time` | Seconds | Temperature measurement time |
|
| `time` | Seconds | Temperature measurement time |
|
||||||
|
| `interval` | Seconds | Time elapsed since last report update on channel |
|
||||||
| `adc` | Volts | AD7172 input |
|
| `adc` | Volts | AD7172 input |
|
||||||
| `sens` | Ohms | Thermistor resistance derived from `adc` |
|
| `sens` | Ohms | Thermistor resistance derived from `adc` |
|
||||||
| `temperature` | Degrees Celsius | Steinhart-Hart conversion result derived from `sens` |
|
| `temperature` | Degrees Celsius | Steinhart-Hart conversion result derived from `sens` |
|
||||||
|
@ -269,7 +270,7 @@ with the following keys.
|
||||||
| `tec_u_meas` | Volts | Measurement of the voltage across the TEC |
|
| `tec_u_meas` | Volts | Measurement of the voltage across the TEC |
|
||||||
| `pid_output` | Amperes | PID control output |
|
| `pid_output` | Amperes | PID control output |
|
||||||
|
|
||||||
Note: With Thermostat v2 and below, the voltage and current readouts `i_tec` and `tec_i` are noisy without the hardware fix shown in [this PR][https://git.m-labs.hk/M-Labs/thermostat/pulls/105].
|
Note: Prior to Thermostat hardware revision v2.2.4, the voltage and current readouts `i_tec` and `tec_i` are noisy without the hardware fix shown in [this PR](https://git.m-labs.hk/M-Labs/thermostat/pulls/105).
|
||||||
|
|
||||||
## PID Tuning
|
## PID Tuning
|
||||||
|
|
||||||
|
|