README: Add PID Output Clamping section
Explains the need of having separate "max_i_pos/output_max" and "max_i_neg/output_min" values; They serve different purposes.
This commit is contained in:
parent
6b8a5f5bb8
commit
3797dce518
16
README.md
16
README.md
@ -239,6 +239,22 @@ of channel 0 to the PID algorithm:
|
||||
output 0 pid
|
||||
```
|
||||
|
||||
### PID output clamping
|
||||
|
||||
It is possible to clamp the PID algorithm output independently of channel output limits. This is desirable when e.g. there is a need to keep the current value above a certain threshold in closed-loop mode.
|
||||
|
||||
Note that the actual output will still be ultimately limited by the `max_i_pos` and `max_i_neg` values, and will not exceed them.
|
||||
|
||||
Set PID maximum output of channel 0 to 1.5 A.
|
||||
```
|
||||
pid 0 output_max 1.5
|
||||
```
|
||||
|
||||
Set PID minimum output of channel 0 to 0.1 A.
|
||||
```
|
||||
pid 0 output_min 0.1
|
||||
```
|
||||
|
||||
## LED indicators
|
||||
|
||||