forked from M-Labs/thermostat
README: Document swap status in report
This commit is contained in:
parent
2c6b8e8186
commit
0940520ded
31
README.md
31
README.md
@ -256,21 +256,22 @@ Use the bare `report` command to obtain a single report. Enable
|
||||
continuous reporting with `report mode on`. Reports are JSON objects
|
||||
with the following keys.
|
||||
|
||||
| Key | Unit | Description |
|
||||
| --- | :---: | --- |
|
||||
| `channel` | Integer | Channel `0`, or `1` |
|
||||
| `time` | Seconds | Temperature measurement time |
|
||||
| `adc` | Volts | AD7172 input |
|
||||
| `sens` | Ohms | Thermistor resistance derived from `adc` |
|
||||
| `temperature` | Degrees Celsius | Steinhart-Hart conversion result derived from `sens` |
|
||||
| `pid_engaged` | Boolean | `true` if in closed-loop mode |
|
||||
| `i_set` | Amperes | TEC output current |
|
||||
| `dac_value` | Volts | AD5680 output derived from `i_set` |
|
||||
| `dac_feedback` | Volts | ADC measurement of the AD5680 output |
|
||||
| `i_tec` | Volts | MAX1968 TEC current monitor |
|
||||
| `tec_i` | Amperes | TEC output current feedback derived from `i_tec` |
|
||||
| `tec_u_meas` | Volts | Measurement of the voltage across the TEC |
|
||||
| `pid_output` | Amperes | PID control output |
|
||||
| Key | Unit | Description |
|
||||
| --- | :---: | --- |
|
||||
| `channel` | Integer | Channel `0`, or `1` |
|
||||
| `time` | Seconds | Temperature measurement time |
|
||||
| `adc` | Volts | AD7172 input |
|
||||
| `sens` | Ohms | Thermistor resistance derived from `adc` |
|
||||
| `temperature` | Degrees Celsius | Steinhart-Hart conversion result derived from `sens` |
|
||||
| `pid_engaged` | Boolean | `true` if in closed-loop mode |
|
||||
| `current_swapped` | Boolean | `true` if TEC current direction is swapped relative to front panel |
|
||||
| `i_set` | Amperes | TEC output current |
|
||||
| `dac_value` | Volts | AD5680 output derived from `i_set` |
|
||||
| `dac_feedback` | Volts | ADC measurement of the AD5680 output |
|
||||
| `i_tec` | Volts | MAX1968 TEC current monitor |
|
||||
| `tec_i` | Amperes | TEC output current feedback derived from `i_tec` |
|
||||
| `tec_u_meas` | Volts | Measurement of the voltage across the TEC |
|
||||
| `pid_output` | Amperes | PID control output |
|
||||
|
||||
Note: With Thermostat v2 and below, the voltage and current readouts `i_tec` and `tec_i` are noisy without the hardware fix shown in [this PR][https://git.m-labs.hk/M-Labs/thermostat/pulls/105].
|
||||
|
||||
|