Add line trigger to shop #227

Merged
sb10q merged 1 commits from architeuthis/web2019:linetrigger into master 2025-10-21 19:23:50 +08:00
Collaborator

#206 (partial)

#206 (partial)
architeuthis added 1 commit 2025-09-18 19:37:18 +08:00
Owner

The panel should say "9-12V AC", definitely not DC.

The panel should say "9-12V AC", definitely not DC.
architeuthis force-pushed linetrigger from 903d82a7ff to edb85c21cd 2025-09-22 04:05:37 +08:00 Compare
Author
Collaborator

image.png

front panel design in the sinara-hw repo seems to be wrong in a variety of ways

![image.png](/attachments/825c78d2-40f2-4a56-807c-48ad2ff3eb21) front panel design in the sinara-hw repo seems to be wrong in a variety of ways
sb10q reviewed 2025-09-23 10:31:32 +08:00
@ -1488,0 +1493,4 @@
image: '/images/shop/graphic-03_Linetrigger.svg',
specs: [
"Low jitter mains trigger box for 50/60 Hz magnetic field noise cancellation",
"115 or 230 V IEC inlet, 50 or 60 Hz, selectable",
Owner

Where is that IEC inlet?

Where is that IEC inlet?
Author
Collaborator

don't think it exists (oops) -- I guess we would ship it together with a wall adapter?

don't think it exists (oops) -- I guess we would ship it together with a wall adapter?
Owner

Yes, and make it clear that the IEC inlet is on that adapter.

Yes, and make it clear that the IEC inlet is on that adapter.
architeuthis marked this conversation as resolved
sb10q reviewed 2025-09-23 10:31:53 +08:00
@ -1488,0 +1498,4 @@
"Separate 9-12 V (RMS) AC input power",
"Dual SMA outputs",
"0 to 5 V trigger signal",
"Can be used with 8452 Stabilizer to drive coils with harmonics of 50/60 Hz with amplitude and phase chosen to cancel out mains-induced noise"
Owner

harmonics?

harmonics?
Author
Collaborator

Taken from the wiki. Have to admit I don't understand it well enough to correct it. I can replace it with an alternate phrasing if someone suggests one.

[Taken from the wiki](https://github.com/sinara-hw/Line_trigger/wiki). Have to admit I don't understand it well enough to correct it. I can replace it with an alternate phrasing if someone suggests one.
architeuthis marked this conversation as resolved
sb10q reviewed 2025-09-23 10:33:28 +08:00
@ -1488,0 +1495,4 @@
"Low jitter mains trigger box for 50/60 Hz magnetic field noise cancellation",
"115 or 230 V IEC inlet, 50 or 60 Hz, selectable",
"Filtered or unfiltered selectable via jumper placement",
"Separate 9-12 V (RMS) AC input power",
Owner

Are you sure it's rms? Considering how this circuit works, the peak voltage is more relevant.

Are you sure it's rms? Considering how this circuit works, the peak voltage is more relevant.
Author
Collaborator
[The issues claim it's RMS](https://github.com/sinara-hw/Line_trigger/issues/5).
Owner

Okay sure but we should aim for a higher standard than those issues/wiki and/or fix the latter.

Okay sure but we should aim for a higher standard than those issues/wiki and/or fix the latter.
Author
Collaborator

Sure. Didn't realize you meant the stat should be replaced. I'll try and figure it out.

Sure. Didn't realize you meant the stat should be replaced. I'll try and figure it out.
architeuthis force-pushed linetrigger from edb85c21cd to 64cb7bddc7 2025-09-23 16:58:00 +08:00 Compare
architeuthis force-pushed linetrigger from 64cb7bddc7 to 863e37f6e3 2025-09-23 17:04:36 +08:00 Compare
sb10q merged commit 37e74f0094 into master 2025-10-21 19:23:50 +08:00
Sign in to join this conversation.
No Reviewers
No Label
2 Participants
Notifications
Due Date
No due date set.
Dependencies

No dependencies set.

Reference: M-Labs/web2019#227
No description provided.