Cleanup fan control #89
13
README.md
13
README.md
|
@ -277,12 +277,11 @@ The thermostat implements a PID control loop for each of the TEC channels, more
|
||||||
|
|
||||||
## Fan control
|
## Fan control
|
||||||
|
|
||||||
Fan control is available for the thermostat revisions with integrated fan system. For this purpose four commands are available:
|
Fan control commands are available for thermostat revisions with an integrated fan system:
|
||||||
1. `fan` - show fan stats: `fan_pwm`, `abs_max_tec_i`, `auto_mode`, `k_a`, `k_b`, `k_c`.
|
1. `fan` - show fan stats: `fan_pwm`, `abs_max_tec_i`, `auto_mode`, `k_a`, `k_b`, `k_c`.
|
||||||
2. `fan auto` - enable auto speed controller mode, which correlates with fan curve `fcurve`.
|
2. `fan auto` - enable auto speed controller mode, where fan speed is controlled by the fan curve `fcurve`.
|
||||||
3. `fan <value>` - set the fan power with the value from `1` to `100` and disable auto mode. There is no way to disable the fan.
|
3. `fan <value>` - set the fan power with the value from `1` to `100` and disable auto mode. There is no way to completely disable the fan.
|
||||||
Please note that power doesn't correlate with the actual speed linearly.
|
Please note that power doesn't correlate with the actual speed linearly.
|
||||||
4. `fcurve <a> <b> <c>` - set coefficients of the controlling curve `a*x^2 + b*x + c`, where `x` is `abs_max_tec_i/MAX_TEC_I`,
|
4. `fcurve <a> <b> <c>` - set coefficients of the controlling curve `a*x^2 + b*x + c`, where `x` is `abs_max_tec_i/MAX_TEC_I`, a normalized value in range [0,1],
|
||||||
i.e. receives values from 0 to 1 linearly tied to the maximum current. The controlling curve should produce values from 0 to 1,
|
i.e. the (linear) proportion of current output capacity used, on the channel with the largest current flow. The controlling curve is also clamped to [0,1].
|
||||||
as below and beyond values would be substituted by 0 and 1 respectively.
|
5. `fcurve default` - restore fan curve coefficients to defaults: `a = 1.0, b = 0.0, c = 0.0`.
|
||||||
5. `fcurve default` - restore fan curve settings to defaults: `a = 1.0, b = 0.0, c = 0.0`.
|
|
||||||
|
|