README: Fix limits section #154

Merged
sb10q merged 1 commits from atse/thermostat:readme-limits-fix into master 2024-11-18 14:01:51 +08:00
Showing only changes of commit 910de3d872 - Show all commits

View File

@ -189,31 +189,30 @@ Testing heat flow direction with a low set current is recommended before install
### Limits ### Limits
Each MAX1968 TEC driver has analog/PWM inputs for setting Each channel has maximum value settings, for setting
output limits. output limits.
Use the `output` command to see current settings and maximum values. Use the `output` command to see them.
| Limit | Unit | Description | | Limit | Unit | Description |
| --- | :---: | --- | | --- | :---: | --- |
| `max_v` | Volts | Maximum voltage | | `max_v` | Volts | Maximum voltage |
| `max_i_pos` | Amperes | Maximum positive current | | `max_i_pos` | Amperes | Maximum positive current |
| `max_i_neg` | Amperes | Maximum negative current | | `max_i_neg` | Amperes | Maximum negative current |
| `i_set` | Amperes | (Not a limit; Open-loop mode) |
Example: set the maximum voltage of channel 0 to 1.5 V. Example: set the maximum voltage of channel 0 to 1.5 V.
``` ```
output 0 max_v 1.5 output 0 max_v 1.5
``` ```
Example: set the maximum negative current of channel 0 to -3 A. Example: set the maximum negative current of channel 0 to -2 A.
``` ```
output 0 max_i_neg 3 output 0 max_i_neg 2
``` ```
Example: set the maximum positive current of channel 1 to 3 A. Example: set the maximum positive current of channel 1 to 2 A.
``` ```
output 0 max_i_pos 3 output 1 max_i_pos 2
``` ```
### Open-loop mode ### Open-loop mode