Noisy measured TEC current (tec_i) #92

Closed
opened 2023-10-09 13:15:21 +08:00 by atse · 2 comments
Contributor

The measured TEC current, tec_i, is noticeably noisier than it is supposed to be.

image (4).png

Most probably related to the constant remeasuring of VREF, used in the current calculation.

The measured TEC current, tec_i, is noticeably noisier than it is supposed to be. ![image (4).png](/attachments/a4a34581-e931-4a7e-a58e-e4083b700e34) Most probably related to the constant remeasuring of VREF, used in the current calculation.
Owner

Most probably related to the constant remeasuring of VREF, used in the current calculation.

I don't think VREF is supposed to fluctuate by ~20% at all.

> Most probably related to the constant remeasuring of VREF, used in the current calculation. I don't think VREF is supposed to fluctuate by ~20% at all.
Owner

Resolved: #105

Resolved: https://git.m-labs.hk/M-Labs/thermostat/pulls/105
Sign in to join this conversation.