README: Correct expected TEC polarity

Adhere to the general convention of TECs cooling down with positive
voltages.
pull/83/head
topquark12 2022-06-06 21:28:30 +08:00 committed by atse
parent 6f0acc73b8
commit 5ba74c6d9b
1 changed files with 1 additions and 1 deletions

View File

@ -184,7 +184,7 @@ postfilter rate can be tuned with the `postfilter` command.
- Connect TEC module device 1 to TEC1- and TEC1+.
- The GND pin is for shielding not for sinking TEC module currents.
When using a TEC module with the Thermostat, the Thermostat expects the thermal load (where the thermistor is located) to heat up with a positive software current set point, and cool down with a negative current set point.
When using a TEC module with the Thermostat, the Thermostat expects the thermal load (where the thermistor is located) to cool down with a positive software current set point, and heat up with a negative current set point.
Testing heat flow direction with a low set current is recommended before installation of the TEC module.